AM1 Alcohol Analyser
The AM1 analyser is used for Clinical & Research applications and offers researchers into clinical alcoholism a very fast method of obtaining results with small animal models. In addition the small sample size (5ul) enables real-time results to be obtained during experiments.
APPLICATION AREAS
- Alcoholism research
OPERATION
After calibration with a standard of the selected metabolite, the single injection of sample is all that is needed to obtain a result and prepare the analyser for the next analysis.
Sample injection via an accurate positive displacement pipette triggers the complete analytical cycle and a hard-copy result is printed within 20 seconds.
ORDER NUMBER
Analox AM1INSTRUMENT SPECIFICATION
| Method | > | Enzymatic oxygen-rate |
| Sensor | > | Clark-type amperometric oxygen electrode |
| Reaction Temp. | > | 30°C |
| Display | > | 32 character backlit LCD |
| Printer | > | 16 column dot matrix, 1 line/sec. |
| Statistics | > | Sequential, giving mean, S.D and C.V. |
| Interface | > | Serial data port, optional Windows software available |
| Power | > | 100-250VAC, 50-60Hz, 12-15VDC, 60VA |
| Dimensions | > | 23cm(width) x 29cm(depth) x 15cm(height) |
| Weight | > | 3.8Kg (Portable, 6Kg) |
Notes:
N/A
ANALYTICAL PERFORMANCE
| Precision | Accuracy (Method Comparisons) | Linearity | ||
|---|---|---|---|---|
| Ethanol (Alcohol) | > | C.V. of 2.5% @ 18.5 mmol/L (ca. 85 mg/dl) (whole blood) |
i) Method comparison for whole blood (neutralised/PCA extract) vs GC: y (Analox) = 1.039x + 0.28 mmol/L, r = 0.991, n = 27 ii) Urine Recovery Data: y (Analox) = 0.981x + 0.19 mmol/L, r = 0.999, n = 17 |
43.0 mmol/L (ca. 200 mg/dl) for 5 µl samples; 86.0 mmol/L (ca. 400 mg/dl) for 2.5 µl samples |
ACCESSORIES
For a full list of ordering info please go to support tab.
| GMRD-023 | > | Heparinised capillaries for plasma alcohol, pack of 500 |
| GMRD-024 | > | Heparinised capillaries for plasma alcohol, pack of 1000 |
| GMRD-110(050) | > | Ethanol standard, 50mg/dl, 4 x 1ml |
| GMRD-110(100) | > | Ethanol standard, 100mg/dl, 4 x 1ml (other values available) |
| GMRD-110(200) | > | Ethanol standard, 200mg/dl, 4 x 1ml |
| GMRD-110(300) | > | Ethanol standard, 300mg/dl, 4 x 1ml |
| GMRD-110D | > | Alcohol quality control, 0.5ml |
| GMRD-110D4 | > | Alcohol quality control,4 x 0.5ml |
| GMRD-113CJ | > | Alcohol reagent kit, 4 x 175ml (4 x 250 analyser cycles) |
Notes:
N/A