AM1 Alcohol Analyser

  • GM7
  • GM7

The AM1 analyser is used for Clinical & Research applications and offers researchers into clinical alcoholism a very fast method of obtaining results with small animal models. In addition the small sample size (5ul) enables real-time results to be obtained during experiments.

APPLICATION AREAS

  • Alcoholism research

OPERATION

After calibration with a standard of the selected metabolite, the single injection of sample is all that is needed to obtain a result and prepare the analyser for the next analysis.

Sample injection via an accurate positive displacement pipette triggers the complete analytical cycle and a hard-copy result is printed within 20 seconds.

ORDER NUMBER

Analox AM1

INSTRUMENT SPECIFICATION

Method > Enzymatic oxygen-rate
Sensor > Clark-type amperometric oxygen electrode
Reaction Temp. > 30°C
Display > 32 character backlit LCD
Printer > 16 column dot matrix, 1 line/sec.
Statistics > Sequential, giving mean, S.D and C.V.
Interface > Serial data port, optional Windows software available
Power > 100-250VAC, 50-60Hz, 12-15VDC, 60VA
Dimensions > 23cm(width) x 29cm(depth) x 15cm(height)
Weight > 3.8Kg (Portable, 6Kg)

Notes:

N/A

ANALYTICAL PERFORMANCE

Precision Accuracy (Method Comparisons) Linearity
Ethanol (Alcohol) > C.V. of 2.5%
@ 18.5 mmol/L
(ca. 85 mg/dl)
(whole blood)
i) Method comparison for whole blood (neutralised/PCA extract) vs GC:
y (Analox) = 1.039x + 0.28 mmol/L,
r = 0.991, n = 27
ii) Urine Recovery Data:
y (Analox) = 0.981x + 0.19 mmol/L,
r = 0.999, n = 17
43.0 mmol/L (ca. 200 mg/dl) for 5 µl samples;
86.0 mmol/L (ca. 400 mg/dl) for 2.5 µl samples

ACCESSORIES

For a full list of ordering info please go to support tab.

GMRD-023 >

Heparinised capillaries for plasma alcohol, pack of 500

GMRD-024 >

Heparinised capillaries for plasma alcohol, pack of 1000

GMRD-110(050) >

Ethanol standard, 50mg/dl, 4 x 1ml

GMRD-110(100) >

Ethanol standard, 100mg/dl, 4 x 1ml (other values available)

GMRD-110(200) >

Ethanol standard, 200mg/dl, 4 x 1ml

GMRD-110(300) >

Ethanol standard, 300mg/dl, 4 x 1ml

GMRD-110D >

Alcohol quality control, 0.5ml

GMRD-110D4 >

Alcohol quality control,4 x 0.5ml

GMRD-113CJ >

Alcohol reagent kit, 4 x 175ml (4 x 250 analyser cycles)

Notes:

N/A