AM3

  • GM7
  • GM7

The AM3 analyser is used in Industrial & Biotechnology analyses for the rapid analysis of ethanol within the range 0 - 10%V/V

APPLICATION AREAS

  • Measurement of ethanol in the beverage industry: beers, lagers, ciders, wines, spirits & alcopops.
  • Waste-line monitoring in distilleries and industrial ethanol plants.
  • Measure of ethanol in OTC consumer health products: cough mixtures, tonics, shampoos.

OPERATION

After calibration with a standard of the selected metabolite, the single injection of sample is all that is needed to obtain a result and prepare the analyser for the next analysis.

Sample injection via an accurate positive displacement pipette triggers the complete analytical cycle and a hard-copy result is printed within 20 seconds.

ORDER NUMBER

Analox AM3

INSTRUMENT SPECIFICATION

Method > Enzymatic oxygen-rate
Sensor > Clark-type amperometric oxygen electrode
Reaction Temp. > 30°C
Display > 32 character backlit LCD
Printer > 16 column dot matrix, 1 line/sec.
Statistics > Sequential, giving mean, S.D and C.V.
Interface > Serial data port, optional Windows software available
Power > 100-250VAC, 50-60Hz, 12-15VDC, 60VA
Dimensions > 23cm(width) x 29cm(depth) x 15cm(height)
Weight > 3.8Kg (Portable, 6Kg)

Notes:

N/A

ANALYTICAL PERFORMANCE

Precision Accuracy (Method Comparisons) Linearity
Ethanol (Alcohol) > 0.01 %V/V, 0.001 %
(selectable)
Beer Comparison vs Reference Distillation (range 2 - 11 %V/V):
y (Analox) = 1.006x + 0.005 %V/V, r = 0.9998
0.1 -10 %V/V

ACCESSORIES

For a full list of ordering info please go to support tab.

GMRD-110(I) >

Alcohol reagent kit, 50ml (70 analyser cycles)

GMRD-110(IJ) >

Alcohol reagent kit, 8 x 50ml (8 x 70 analyser cycles)

GMRD-110X >

A range of alcohol standards at specified (x%V/V) concentrations

GMRD-112 >

Alcohol reagent kit, 4 x 175ml (4 x 250 analyser cycles)

GMRD-117A >

Dilution tubes, pack of 500

GMRD-119 >

Degassing tubes, pack of 100

GMRD-120 >

Rack for dilution tubes

SA4 >

Vortex mixer unit

Notes:

N/A