The GM10 analyser is used in Industrial & Biotechnology analyses for the rapid analysis of sugars (ß-D-glucose–Dextrose, sucrose and lactose) in food & beverage, biopharmaceutical and industrial process control applications.

APPLICATION AREAS

Suitable for the analysis of sugars in
the following applications:

  • Food & beverages
  • Pharmaceutical
  • Biotechnology
  • Industrial process control
  • Cell culture
  • Fermentation and bioprocess monitoring

OPERATION

After calibration with a standard of the selected metabolite, the single injection of sample is all that is needed to obtain a result and prepare the analyser for the next analysis.

Sample injection via an accurate positive displacement pipette triggers the complete analytical cycle and a hard-copy result is printed within 20 seconds.

ORDER NUMBER

Analox GM10

INSTRUMENT SPECIFICATION

Method > Enzymatic oxygen-rate
Sensor > Clark-type amperometric oxygen electrode
Reaction Temp. > 30°C
Display > 32 character backlit LCD
Printer > 16 column dot matrix, 1 line/sec.
Statistics > Sequential, giving mean, S.D and C.V.
Interface > Serial data port, optional Windows software available
Power > 100-250VAC, 50-60Hz, 12-15VDC, 60VA
Dimensions > 23cm(width) x 29cm(depth) x 15cm(height)
Weight > 3.8Kg (Portable, 6Kg)

Notes:

N/A

ANALYTICAL PERFORMANCE

Precision Accuracy (Method Comparisons) Linearity
Glucose > 0.01 g/L @2g/L N/A 0.05 - 20 %W/V
Sucrose > 0.01 %W/V N/A 0.5 - 40 %W/V
Lactose > 0.01 %W/V N/A 0.5 - 40%W/V

ACCESSORIES

For a full list of ordering info please go to support tab.

GMRD-002A >

Glucose oxidase reagent, 360 analyser cycles

GMRD-015 >

Glucose standard, 1.0%W/V, 30ml

GMRD-016 >

Glucose standard, 5.0%W/V, 30ml

GMRD-017 >

Glucose standard, 8.0%W/V, 30ml

GMRD-018 >

Glucose standard, 20.0%W/V, 30ml

GMRD-095S >

Lactose reagent kit, 50-100 tests, including glucose reagent & standard

GMRD-105S >

Sucrose reagent kit, 50-100 tests, including glucose reagent & standard

Notes:

N/A