GM7
The GM7 analyser is used for Clinical & Research applications and enables researchers to obtain fast, accurate results for key analytes in clinical research studies. Its unique analysis menu includes programs for acetoacetate, alcohol, ammonia, cholesterol, creatinine, glucose, glutamine, glycerol, 3-Hydroxybutyrate, lactate, pyruvate, triglycerides, urea, and uric acid
APPLICATION AREAS
- Diabetes Research – Clamp studies
- Clinical Research
- Metabolic studies
- Biochemical Research
- Sports Medicine
OPERATION
After calibration with a standard of the selected metabolite, the single injection of sample is all that is needed to obtain a result and prepare the analyser for the next analysis.
Sample injection via an accurate positive displacement pipette triggers the complete analytical cycle and a hard-copy result is printed within 20 seconds.
ORDER NUMBER
Analox GM7INSTRUMENT SPECIFICATION
| Method | > | Enzymatic oxygen-rate |
| Sensor | > | Clark-type amperometric oxygen electrode |
| Reaction Temp. | > | 30°C |
| Display | > | 32 character backlit LCD |
| Printer | > | 16 column dot matrix, 1 line/sec. |
| Statistics | > | Sequential, giving mean, S.D and C.V. |
| Interface | > | Serial data port, optional Windows software available |
| Power | > | 100-250VAC, 50-60Hz, 12-15VDC, 60VA |
| Dimensions | > | 27cm(width) x 35cm(depth) x 25cm(height) |
| Weight | > | 5.8Kg |
Notes:
Example data below, please see Analyte Technical Bulletins for full specifications.
ANALYTICAL PERFORMANCE
| Precision | Accuracy (Method Comparisons) | Linearity | ||
|---|---|---|---|---|
| Glucose | > | C.V. of 1.0% @ 5mmol/L (plasma) |
Method comparison vs Hexokinase: y (Analox) = 0.9836x - 0.1335 mmol/L, r = 0.09991, n = 147 |
30.0 mmol/L (540 mg/dl) for 10 µl samples |
| Lactate | > | C.V. of 2.0% @ 2.5 mmol/L |
Method comparison for whole blood vs YSI 23L: y (Analox) = 0.98x + 0.055 mmol/L, r = 0.9991, n = 56 |
10 mmol/L (ca. 90 mg/dl) for 7 µl samples |
| Alcohol | > | C.V of 2.5% @ 18.5 mmol/L (ca. 85mg/dl) (whole blood) |
Method comparison for whole blood (neutralised PCA extract) vs GC: y (Analox) = 1.039x = 0.28 mmol/L, r = 0.991, n = 27 |
43.0 mmol/L (ca. 200 mg/dl) for 5 µl samples |
| Urea | > | C.V. of 1.1 % @ 14.0 mmol/L (ca. 85 mg/dl) |
i) Method comparison vs Beckman urease conductimetric proceedure: y(Analox) = 0.993x + 0.15 mmol/L, r = 0.998, n = 100 ii) Method comparison vs Urease/Berthelot method: y(Analox) = 1.01x - 0.108 mmol/L, r = 0.999, n = 93 |
45 mmol/L (ca. 270 mg/dl) |
ACCESSORIES
For a full list of ordering info please go to support tab.
| GMRD-100/103/102 | > | Lactate reagent for lysed whole blood - 100, 250 and 1000 analyser cycles respectively, including 8.0mmol/L standard |
| GMRD-002A | > | Glucose oxidase reagent, 360 analyser cycles |
| GMRD-011 | > | Glucose standard, 8mmol/L, 30ml (other values available) |
| GMRD-021 | > | Uric acid reagent kit, 70 analyser cycles incl. 0.6mmol/L standard |
| GMRD-084 | > | Cholesterol reagent kit, 50 tests, incl. 5.2mmol/L standard |
| GMRD-090/093/092 | > | Lactate reagent for intact whole blood - 100, 250 and 1000 analyser cycles respectively, including 8.0mmol/L standard |
| GMRD-110(100) | > | Ethanol standard, 100mg/dl, 4 x 1ml (other values available) |
| GMRD-113 | > | Alcohol reagent kit, 70 analyser cycles |
| GMRD-130 | > | 3-hydroxybutyrate reagent kit, 20-45 tests, incl. standard |
| GMRD-140 | > | Pyruvate reagent kit, 20-40 tests, incl. standard |
| GMRD-150 | > | Glutamine reagent kit, 50 tests, incl. standard |
| GMRD-162 | > | Ammonia reagent kit, 20-40 tests, incl. standard |
| GMRD-170 | > | Creatinine reagent kit, 40-80 tests, incl. standard |
| GMRD-177 | > | Glycerol reagent kit, 140 analyser cycles, incl. standard |
| GMRD-180 | > | Acetoacetate reagent kit, 20-40 tests, incl. standard |
| GMRD-195 | > | Triglycerides reagent kit, 140 analyser cycles, incl. standard |