GM9D Glucose Direct

  • GM7
  • GM7

The GM9D analyser is ideal for glucose clamping studies, giving a printed whole blood result from 35 seconds without the requirement for any specialised collection system.

APPLICATION AREAS

  • Diabetes Research – glucose clamp studies

OPERATION

After calibration with a standard of the selected metabolite, the single injection of sample is all that is needed to obtain a result and prepare the analyser for the next analysis.

Sample injection via an accurate positive displacement pipette triggers the complete analytical cycle and a hard-copy result is printed within 20 seconds.

ORDER NUMBER

Analox GM9D

INSTRUMENT SPECIFICATION

Method > Enzymatic oxygen-rate
Sensor > Clark-type amperometric oxygen electrode
Reaction Temp. > 30°C
Display > 32 character backlit LCD
Printer > 16 column dot matrix, 1 line/sec.
Statistics > Sequential, giving mean, S.D and C.V.
Interface > Serial data port, optional Windows software available
Power > 100-250VAC, 50-60Hz, 12-15VDC, 60VA
Dimensions > 23cm(width) x 29cm(depth) x 15cm(height)
Weight > 3.8Kg (Portable, 6Kg)

Notes:

N/A

ANALYTICAL PERFORMANCE

Precision Accuracy (Method Comparisons) Linearity
Glucose > C.V. of 2 % @ 5 mmol/L
(whole blood)
N/A
30.0 mmol/L (540 mg/dl)
for 10µl samples;
50.0 mmol/L (900 mg/dl)
5µl samples

ACCESSORIES

For a full list of ordering info please go to support tab.

GMRD-009 >

Glucose standard, 2.5mmol/L, 30ml

GMRD-010 >

Glucose standard, 5.0mmol/L, 30ml

GMRD-011 >

Glucose standard, 8mmol/L, 30ml (other values available)

GMRD-012 >

Glucose standard, 25.0mmol/L, 30ml

GMRD-020 >

Reagent kit for GM9D analyser, 250 tests, inc. 8mmol/L standard

Notes:

N/A