In addition, the GM9 offers whole blood glucose analysis when the appropriate Analox collection system is used. A fully portable model (P-GM9) is also available. Analytical Principle: The
GM9 measures the rate of oxygen uptake in the reaction between
sample glucose
and glucose oxidase (GOD). Under the assay conditions, this is directly proportional
to the glucose concentration:
Operation: The single action of injecting a sample is all you need to obtain a result and prepare the analyser for the next analysis. Sample injection via a very accurate positive displacement pipette also triggers the complete analytical cycle. A hard-copy result is obtained on the integral printer within 20 seconds of sample introduction. The RS232 port facilitates interface with your PC database. Samples can be whole blood, plasma, serum or other biological fluids. Whole blood samples are collected in Analox capillaries or tubes containing heparin, fluoride and nitrite. Analyses are menu-driven via the 32-character display which guides the operator through the complete procedure. Subsidiary menus are reached via YES and NO buttons which enable the user to change calibration values, optimise operational modes, utilise special functions and perform statistical data analysis. The display also provides self-test diagnostics in relation to sensor electrode status and fluid movements. |
Analytical Performance |
||
Linearity |
0 - 30mmol/L using 10µl sample, 0 - 50mmol/L using 5µl sample |
|
Accuracy (Recovery) |
Plasma,
mean 100.92% (99.2-106.7) |
Whole
blood, mean 96.7% (89.1-100.2) |
Accuracy
(Method Comparison) |
Plasma,
vs GOD-PAP AAII |
Whole
blood, vs GOD-PAP AAII |
Precision |
Plasma,
n = 20 |
Whole
blood, n = 40 |
Sensitivity |
0.1mg/dl, 0.01mmol/L (selectable) |
|
Instrument
Specification: |
|
Method |
Enzymatic oxygen-rate |
Ordering Information: |
|
GM9 |
Glucose analyser, mains version |
GMRD-053/054 |
Blood collection capillaries for intact whole
blood glucose, |
GMRD-050 |
Record cards for collection capillaries, pack
of 40 (1000 samples) |
GMRD-070/072 |
Heparinised capillary/microcentrifuge
tube, lysing, for 'total' blood |
GMRD-047 |
Lysing tubes for 'total' blood glucose, 0.2ml
sample, pack of 500 |
|
|