The integral rechargeable battery allows the analyser to be used at track or poolside and the optional all-weather carrying case enables the P-GL5 to be operated in virtually any environment. A mains version, the Model GL5, is also available. Analytical Principle: In oxidase enzyme reactions, the analyser measures the rate of oxygen uptake and under appropriate controlled conditions, this is directly proportional to the concentration of the analyte. Example:
Operation: For each assay, the simple injection of a sample is all that is needed to obtain a result and prepare the analyser for the next analysis. Sample injection via an accurate positive displacement pipette triggers the complete analytical cycle and a hard-copy result is then printed within 20 seconds. The analyser is first calibrated with a standard of the metabolite in question. Samples can be whole blood (for key assays such as lactate and glucose), plasma, serum or other biological fluids Analyses are menu-driven via the 32-character display which guides the operator through the complete procedure. Subsidiary menus are reached via YES and NO buttons which enables the user to change analysis type, optimise operational modes, utilise special functions and perform statistical data analysis. The display also provides self-test diagnostics in relation to electrode status and fluid movements. Changeover between analytes is quickly and simply effected and all fluid pathways inside the analyser can be rapidly sterilised without compromising performance. |
Analytical Performance |
|||
|
Accuracy (Method Comparisons) |
Linearity |
Precision |
Glucose |
y (Analox) =
0.98 (HK) - 2.5mg/dl |
0 -540mg/dl |
CV = 1% |
Lactate |
y (Analox) =
0.98 (YSI) + 0.8mg/dl |
0 - 90mg/dl |
CV = 0.5 - 1.0% |
Alcohol |
y (Analox) =
0.97 (True Val.) + 2.1mg/dl |
0 - 300mg/dl |
CV = 2.5% |
Cholesterol |
y (Analox) =1.06
(Man.Enz) - 7.7mg/dl |
0 - 400mg/dl |
CV = 1% |
Urate |
y (Analox) =
1.02 (AAII) +0.06mg/dl |
0 - 13.5mg/dl |
CV = 2% |
Instrument
Specification: |
|
Method |
Enzymatic oxygen-rate |
Power |
100-250VAC, 50-60Hz, 12-15VDC,
60VA, plus 5VA |
Dimensions |
23cm(width) x 29cm(depth)
x 15cm(height) |
Ordering Information: |
|
P-GL5 Portable |
Portable analyser, rechargeable battery/mains |
GMRD-090/093/092 |
Lactate reagent for intact whole blood - 100,
250 and 1000 analyser |
GMRD-100/103/102 |
Lactate reagent for lysed whole blood - 100,
250 and 1000 analyser |
GMRD-113 |
Alcohol reagent kit, 70 analyser cycles |
Specialised blood collection systems and accessories are available – please contact Analox for details |
|