The Model P-GM7 is ideal for routine training monitoring, sports science teaching courses and research projects. A mains version, Model GM7, is also available. Analytical Principle: In oxidase enzyme reactions, the analyser measures the rate of oxygen uptake and under appropriate controlled conditions, this is directly proportional to analyte concentration. Example:
Substrates for dehydrogenase enzymes are measured using two stage reactions in which the oxidation of NADH (produced or consumed) is monitored. Example:
Under the conditions of the assay, the rate of oxidation of excess NADH by peroxidase (POD) is inversely proportional to ammonia concentration.
Operation: For many assays, the simple injection of a sample is all that is needed to obtain a result and prepare the analyser for the next analysis. Sample injection via an accurate positive displacement pipette triggers the complete analytical cycle and a hard-copy result is printed within 20 seconds. The analyser is first calibrated with a standard of the metabolite in question. Samples can be whole blood (e.g. key assays such as lactate, glucose, urea), plasma, serum or other biological fluids. Analyses are menu-driven via the 32-character display which guides the operator through the complete procedure. Subsidiary menus are reached via the keypad which enables the user to change analysis type, optimise operational modes, utilise special functions and perform statistical data analysis. The display also provides self-test diagnostics in relation to electrode status and fluid movements. Changeover between analytes is quickly and simply effected and all fluid pathways inside the analyser can be rapidly sterilised without compromising performance. |
Analytical Performance |
|||
|
Glucose |
Lactate |
Alcohol |
Precision |
Plasma; n=20,
CV=1.4% |
Plasma; n=30,
CV=0.6% |
Plasma; CV=1-2% |
Accuracy - Recovery |
Plasma; mean
100.9% |
Plasma; mean
99.2% |
Spiked serum; |
Accuracy - Method |
Plasma vs Hexokinase |
Plasma vs Boehringer
UV, |
Whole blood
vs GC, |
Linearity |
0 - 30mmol/L
(540mg/dl) |
0 - 10mmol/L
(90mg/dl) |
0 - 200mg/dl
(43mmol/L) |
Instrument Specification: |
|
Method |
Enzymatic oxygen-rate |
Statistical programmes |
a)Sequential, giving mean, S.D
and C.V. |
Interface |
Serial data port, optional Windows software
available |
Ordering Information: |
|
P-GM7 MicroStat |
Portable analyser, rechargeable battery/mains |
GMRD-090/093/092 |
Lactate reagent for intact
whole blood - 100, 250 and 1000 analyser |
GMRD-100/103/102 |
Lactate reagent for lysed whole
blood - 100, 250 and 1000 analyser |
GMRD-140 |
Pyruvate reagent kit, 20-40
tests, incl. standard |
Specialised blood collection systems and accessories are available – please contact Analox for details |
|