The integral rechargeable battery allows the analyser to be used at track or poolside and an optional all-weather carrying case enables the P-LM5 to be operated in virtually any environment. A mains version, the Model LM5 Champion, is also available. Analytical Principle: With lactate oxidase enzyme, the analyser measures the rate of oxygen uptake and, under appropriate controlled conditions, this is directly proportional to the lactate concentration. Example:
Operation: The single action of injecting a sample is all you need to obtain a result and prepare the analyser for the next analysis. Sample injection via a very accurate positive displacement pipette also triggers the complete analytical cycle. A hard-copy result is then obtained within 20 seconds of sample introduction. The analyser is first calibrated with a known lactate standard. Samples can be whole blood, plasma, serum or other biological fluids. Whole blood samples are collected in Analox capillaries or tubes containing heparin, fluoride and nitrite. Analyses are menu-driven via the 32-character display which guides the operator through the complete procedure. Subsidiary menus are reached via YES and NO buttons which enable the user to change calibration values, optimise operational modes, utilise special functions and perform statistical data analysis. The display also provides self-test diagnostics in relation to sensor electrode status and fluid movements. |
Analytical Performance |
||
Linearity |
0-10mmol/L using 5µl sample, 0-20mmol/L using 2.5µl sample |
|
Accuracy (Recovery) |
Plasma, mean
99.25% (96-104%) |
Whole blood,
mean 97% (88-104%) |
Accuracy |
Plasma, vs Boehringer
UV |
Whole blood,
(non lysed), vs YSI23L |
Precision |
Plasma, n =
30 |
Whole blood,
n = 22 |
Sensitivity |
0.1mg/dl, 0.01%mmol/L (selectable) |
|
Instrument
Specification: |
|
Method |
Enzymatic oxygen-rate |
Ordering Information: |
|
P-LM5 Portable |
Portable analyser, rechargeable battery/mains |
GMRD-090/093/092 |
Lactate reagent for intact whole blood - 100,
250 and 1000 analyser |
GMRD-100/103/102 |
Lactate reagent for lysed whole blood - 100,
250 and 1000 analyser |
GMRD-074 |
Lactate quality control, 4 x 0.5ml |
Specialised blood collection systems and accessories are available – please contact Analox for details |
|