Technical Summary: Clinical
Alcohol
Analysis
(Bulletin ref. TB10914E) |
|
|
A rapid
high performance plasma alcohol analysis using Alcohol Oxidase (AOD). |
|
|
Name: |
Alcohol |
|
Codes: |
GMRD-113
Alcohol reagent kit, 50ml (70 analyser cycles) |
|
|
GMRD-110(050)
Ethanol standard, 50mg/dl, 4 x 1ml |
|
|
GMRD-110D
Alcohol quality control, 0.5ml |
|
Instruments: |
All
GM7/AM1/GL5 Series analysers |
|
Principle: |
In
the
presence of molecular oxygen, ethanol is oxidised by the enzyme
alcohol oxidase to acetaldehyde and hydrogen peroxide: |
|
|
|
|
|
|
|
Samples: |
Plasma, serum,
urine, precipitated whole blood (neutralised) |
|
Precision: |
Typical within-run C.V.
= 2.5 % @ 85 mg/dl (whole blood) |
|
Accuracy: |
i) Serum Recovery Data: |
|
Reagent Stability: |
Shelf-life 6 months unopened when stored at 0
- 5°C. Reconstituted AOD/buffer reagent,
ca 5 days if stored at 0 - 5°C between use.
Fresh aliquots may be frozen for several weeks at -20°C
or lower to extend life. |
|
Note: |
Alcohol oxidase is non-specific towards ethanol. This assay will also detect other low molecular weight alcohols, e.g. methanol. |