Technical Summary: Glucose Analysis (Bulletin ref. TB108932B)
 

A very fast enzymatic assay for glucose in plasma, serum or whole blood via Analox
collection systems
 

Name:
Code:
Size:

Glucose
GMRD-002A
250ml (360 analyser cycles)

Instruments:
 

All GM7/GM9/GL5 Series analysers
 

Principle:

In the presence of molecular oxygen, glucose is oxidised by the enzyme glucose oxidase to gluconic acid and hydrogen peroxide:
 

 


Under the conditions of the assay, oxygen consumption is directly proportional to glucose concentration.
 

Samples:

Plasma, serum, whole-blood collected in Analox tubes or capillaries, C.S.F., urine (suitably diluted) and other aqueous media.
 

Sample volume:
Analysis time

10 µl (variable 3.5 - 25 µl)
20 seconds
 

Linearity:

30 mmol/L (10 µl sample); 540 mg/dl
50 mmol/L (5 µl sample); 900 mg/dl
 

Precision:

Plasma, n = 20, CV=1.4 % at 10 mmol/L
Whole-blood, n = 40, CV = 0.85 % at 12 mmol/L
 

Accuracy:

i) Whole-blood comparison vs GOD PAP AAII
y (Analox) = 1.02x - 0.26 mmol/L, r = 0.998; n = 100
ii) Serum comparison vs Hexokinase
y (Analox) = 0.98x - 0.14 mmol/L, r = 0.999; n = 156
 

Reagent Stability:

Shelf-life 15 months stored at 0 - 5°C.


Note:


Glucose standards at 5.0, 8.0 and 25.0 mmol/L (30 ml) are available and ordered as required. Other values at 30.0 and 50.0 mmol/L can be supplied to special order.